Fume exhaust hood In Chennai

Pure Air System is a Prominent & Leading Manufacturer Fume Exhaust Hood In Chennai, we offer fume exhaust hoods, bypass fume hood and fume hoods at best price with quality. In accordance with certified standards, the manufacturing of the exhaust hoods we offer makes use of superior materials and polished granite or glazed ceramic tiles.

Pureair” Fume Exhaust Hood ( Constant volume type) is fitted with a movable front sash and an interior baffle. The sash may be fully opened to accommodate apparatus within the fume hood, while baffle directs exhaust air along specific flow pattern within the enclosure. The constant volume type generally operates at constant exhaust volume with the total exhaust air entering the hood through the sash opening. Closing the sash increases the speed of the air through the sash opening, so that high velocities are to be expected with the sash in the near closed positions.

ProductPureair Fume Exhaust Hood – Constant Volume Type
DimensionsAs per drawings enclosed.
MOCGalvanized Iron with PU coated / Stainless Steel – Satin Finish
Front Sash HeightMaximum up to 700mm
Exhaust CollarMinimum 6” dia PVC pipe
Motor Blower AssemblyInduction Motor with Centrifugal Blower Statically and Dynamically balanced with Suitable Rating and Size.
MountingUnique Spring Suspension Systems to take care of Vibration and Noise.
Working TableGranite Stone / FRP Moulded Top / SS 304 Stainless Steel Top
Front SashClear Glass Side Panels / Clear Acrylic Panel / UV Protected Polycarbonate Panel
Airflow MonitorTEK UK make Face Velocity Indicator with Alarm. (Optional)
LightsFluorescent Tube with Milky White Diffuser.
Other Electricals3 Pin Socket, 15 AMP – 1 No. – On / Off Switches with Indicator.
Performance
Standard ReferenceASHRAE-110 and EN14175-3
Face Velocity0.45 ± 0.05 m/s or 90 FPM ± 10 FPM
Illumination Of LightMinimum 800 Lux on work table
Noise LevelLess than 67 dB on Scale 'A' when ambient sound level is not greater than 57 dB on Scale 'A'
Vibration LevelMinimum

Bypass type Fume Exhaust Hood

Pureair” make Fume Hood (By-Pass Type) are designed in such a manner that closing the front sash causes a re- distribution of air entering the hood through the sash opening eliminating high velocity air streams in the near closed position of the sash. The bypass opening above the sash minimizes fluctuations in face velocity as the sash is raised or lowered. The bypass hood is more desirable for use with fragile apparatus and instrumentation. Since the face velocity does not reach levels which might be detrimental to experimentation

Product Pureair Fume Exhaust Hood – Bypass Type
Dimensions As per drawings enclosed.
MOC Galvanized Iron with PU coated / Stainless Steel – Satin Finish / Wooden laminated
Front Sash Height Maximum up to 700mm
Exhaust Collar Minimum 6” dia PVC pipe
Motor Blower Assembly Induction Motor with Centrifugal Blower Statically and Dynamically balanced with Suitable Rating and Size.
Mounting Unique Spring Suspension Systems to take care of Vibration and Noise.
Working Table Granite Stone / FRP Moulded Top / SS 304 Stainless Steel Top
Front Sash Clear Glass Side Panels / Clear Acrylic Panel / UV Protected Polycarbonate Panel
Airflow Monitor TEK UK make Face Velocity Indicator with Alarm.
Lights Fluorescent Tube with Milky White Diffuser.
Other Electricals 3 Pin Socket, 15 AMP – 1 No. – On / Off Switches with Indicator.
Performance
Standard Reference ASHRAE-110 and EN14175-3
Face Velocity 0.45 ± 0.05 m/s or 90 FPM ± 10 FPM
Illumination Of Light Minimum 800 Lux on work table
Noise Level Less than 67 dB on Scale ‘A’ when ambient sound level is not greater than 57 dB on Scale ‘A’
Vibration Level Minimum

For More Enquiry regarding Bio safety Cabinets & Price, Please Mail to
image